Urgent site urgencyFDA Class I

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no

Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Agency
FDA Device
Recall date
May 15, 2026
Status
Ongoing
Company
Medline Industries, LP

1. What product was recalled?

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no

Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.

Quantity: 3

Code or lot text: Lot EP250203 Item no. 401206RH - UDI-DI 10197344043767; Item no. 401305RH - UDI-DI 10197344043637.

LOT
Item no. 401305RH - UDI-DI 10197344043637.
LOT
Lot EP250203 Item no. 401206RH - UDI-DI 10197344043767
RECALL-NUMBER
Z-2615-2026

2. Is my exact product affected?

Compare every printed identifier. A matching value can indicate that your product is included.

3. What is the hazard?

Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

4. What should I do now?

Follow the FDA recall notice and contact the recalling company for current instructions.

5. Where was it sold or distributed?

Nationwide

View source evidence for state distribution
  • Texas: US Nationwide distribution in the state of TX.

6. Were injuries or illnesses reported?

No separate injury or illness statement was present in the normalized record. Check the official source for the latest report.

7. What remedy is available?

Follow the FDA recall notice and contact the recalling company for current instructions.

8. Is the recall active or closed?

The source status is Ongoing. The Recall Guy's urgency label is urgent, which is a separate deterministic site label.

9. When was the information checked?

Last checked .

  1. created

10. Where is the official source?

Open the official FDA Device source

The Recall Guy normalizes public source data for easier checking. Source agencies remain authoritative. Confirm identifiers and instructions at the official source.