Omnipod (Eros) Pod
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
- Agency
- FDA Device
- Recall date
- May 20, 2026
- Status
- Ongoing
- Company
- Insulet Corporation
1. What product was recalled?
Omnipod (Eros) Pod
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
Quantity: 289,046 units
Code or lot text: 1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514.
- LOT
- 1. Model/REF Number: POD-ZXP425 (10 Pack)
- LOT
- L71480
- LOT
- L72509
- LOT
- L72510
- LOT
- L72511
- LOT
- L72513
- LOT
- L72514.
- LOT
- L72521. 3. Model/REF Number: ZXP425 (10 Pack)
- LOT
- POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack)
- LOT
- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515
- LOT
- ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516
- RECALL-NUMBER
- Z-2605-2026
2. Is my exact product affected?
Compare every printed identifier. A matching value can indicate that your product is included.
3. What is the hazard?
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
4. What should I do now?
Follow the FDA recall notice and contact the recalling company for current instructions.
5. Where was it sold or distributed?
Nationwide
6. Were injuries or illnesses reported?
No separate injury or illness statement was present in the normalized record. Check the official source for the latest report.
7. What remedy is available?
Follow the FDA recall notice and contact the recalling company for current instructions.
8. Is the recall active or closed?
The source status is Ongoing. The Recall Guy's urgency label is urgent, which is a separate deterministic site label.
9. When was the information checked?
Last checked .
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
- updated
10. Where is the official source?
Open the official FDA Device source
The Recall Guy normalizes public source data for easier checking. Source agencies remain authoritative. Confirm identifiers and instructions at the official source.