Urgent site urgencyFDA Class I

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

Agency
FDA Device
Recall date
May 6, 2026
Status
Ongoing
Company
Fresenius Kabi USA, LLC

1. What product was recalled?

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Quantity: 18,444

Code or lot text: Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:

LOT
All Serial/Lot Numbers:
LOT
Model Number: LVP-0004
LOT
UDI-DI: 00811505030320
RECALL-NUMBER
Z-2224-2026

2. Is my exact product affected?

Compare every printed identifier. A matching value can indicate that your product is included.

3. What is the hazard?

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

4. What should I do now?

Follow the FDA recall notice and contact the recalling company for current instructions.

5. Where was it sold or distributed?

California, Colorado, Florida, Georgia, Idaho, Illinois, Maryland, Michigan, Minnesota, Missouri, Mississippi, Nebraska, New Jersey, Nevada, Oklahoma, Oregon, South Carolina, Texas, Virginia, Washington, Wisconsin

View source evidence for state distribution
  • California: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Colorado: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Florida: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Georgia: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Idaho: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Illinois: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Maryland: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Michigan: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Minnesota: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Missouri: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Mississippi: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Nebraska: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • New Jersey: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Nevada: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Oklahoma: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Oregon: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • South Carolina: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Texas: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Virginia: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Washington: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
  • Wisconsin: Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;

6. Were injuries or illnesses reported?

No separate injury or illness statement was present in the normalized record. Check the official source for the latest report.

7. What remedy is available?

Follow the FDA recall notice and contact the recalling company for current instructions.

8. Is the recall active or closed?

The source status is Ongoing. The Recall Guy's urgency label is urgent, which is a separate deterministic site label.

9. When was the information checked?

Last checked .

  1. created

10. Where is the official source?

Open the official FDA Device source

The Recall Guy normalizes public source data for easier checking. Source agencies remain authoritative. Confirm identifiers and instructions at the official source.